XAR-OZ Ozone Gas Detector (0-20 ppm)

Products > Ozone > XAR-OZ Ozone Gas Detector (0-20 ppm)

XAR-OZ Ozone Gas Detector (0-20 ppm)

XAR-OZ is a precise gas detector for monitoring ozone (O3) in the workplace. It has been designed to notify of the presence of ozone (O3) gas.

XAR-OZ Ozone Gas Detector


Santacary XAR-OZ is a precise gas detector for monitoring ozone (O3) in the workplace. It has been designed to notify of the presence of ozone (O3) gas. XAR-OZ portable gas detector has a wide range of applications in industrial, business, home or R&D and other fields. It is useful for detection of ozone residue after disinfection.


Features

Portable O3 gas detector

O3 gas detector range: 0 ~ 20 PPM. Resolution: 0.1 PPM

Audible alarm

Support zero and span calibration

Two points of instantaneous alarm

Trend chart display showing the past readings for O3

With temperature and humidity measurement

Only two buttons and easy to operate

Four AA Alkaline Batteries

 

Note:  XAR-OZ should be used in a strong convection air environment.

 

unit Description

Device

1.        Liquid crystal display (LCD)

2.        Power button

3.        Function button

4.        Battery compartment cover

5.        Air sampling ports

 image.png


DISPLAYs

Normal Display

                                              image.png

 

1.        Alarm Status (None/Low Alarm/High Alarm)

2.        Battery Level

3.        O3 concentration

4.        O3 concentration unit (PPM)

5.        Air Temperature

6.        % Relative Humidity


Alarm

Low Alarm and High Alarm

XAR-OZ has two alarm set points: High Alarm (O3: 10.0 PPM) and Low Alarm (O3: 5.0 PPM). These set points are factory set and cannot be changed. XAR-OZ is equipped with audio alarms to alert you when the ambient gas concentration exceeds one of the two alarm set points. When O3 value exceeds the defined high alarm set point (10.0 PPM), the audio alarm will sound at 3 beeps/sec. When O3 value exceeds the defined low alarm set point (5.0 PPM) but less than the defined high alarm set point (10.0 PPM), the audio alarm will sound at 2 beeps/sec.

 

Factory Alarm Set points

Gas

Low

High

O3

5.0 PPM

10.0 PPM

 image.png


Specifications

O3 Sensor Specification:

Sensor

Electrochemical sensor

Sample Method

Diffusion

Measurement Range

0~20 PPM

Resolution

0.1 PPM

Repeatability

<±5% of signal

Accuracy

±5%FS

Warm-up time

<3 mins

Response time

< 120 seconds (diffusion)

Recovery time

< 120 seconds (diffusion)

Service life

2 years (in air)

 

Temperature Specification

Temperature Range

-10.0~60.0°C (14~140°F) display

Display Resolution

0.1°C (0.1°F)

Display Options

°C/°F switchable

Accuracy

±0.5°C (±0.9°F)

Response Time

5~30 seconds (device must equilibrate with   environment)

 

RH Specification

Measurement   Range

0.0~99.9%RH

Display   Resolution

1%RH

Accuracy

±4.5%RH

Response Time

<8 seconds   for 63% of step change

 

General Specification

Operating

-10°C to 50°C (14°F to 122°F), 15~90%   RH non-condensing

Storage

-10°C to 60°C (14°F to 140°F), <99%   RH non-condensing

Power Supply

Four AA Alkaline Batteries

Dimensions

75x165x25mm (2.95x6.49x0.98”)

Weight

125 grams (4.41 oz.)   without batteries

 

Out of range of operating conditions will impact the accurate of O3 measurement.

 

Warranty

The XAR-OZ is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. This warranty covers normal operation and does not cover misuse, abuse, alteration, neglect, improper maintenance. Proof of purchase is required for warranty. Warranty is void if the detector has been opened.