H2S Gas Detector XAR-HP

Products > Gas Detectors > H2S Gas Detector XAR-HP

H2S Gas Detector XAR-HP

H2S Gas Detector XAR-HP

Features

Portable H2S gas detector

H2S gas detector range: 0 ~ 500 PPM. Resolution: 1.0 PPM

Using high precision 3-electrodes electrochemical H2S sensor

Support auto zero and span calibration

One press to restore factory setting, free from the bother of mis-operation

Audible alarm

Two set points of instantaneous alarm

Trend chart display showing the past readings for H2S

With temperature and humidity measurement

Only two buttons and easy to operate

Sensor is placed at the front end of gooseneck for quick measurement

Four AA alkaline batteries

 

Theory of operation

The Santacary XAR-HP H2S Gas Detector uses 3-electrode electrochemical technology, operating by the diffusion principle, for determining the concentration of H2S in air samples. In diffusion mode, the atmosphere reaches the sensor in the front end of gooseneck by diffusing. Normal air movements are enough to carry the sample to the sensor.

 

Device

1.        Sensor at the front end of gooseneck

2.        Liquid crystal display (LCD)

3.        Power button

4.        Function button

5.        Battery compartment cover


Alarm

Low Alarm and High Alarm

XAR-HP has two alarm set points: High Alarm (H2S: 15 PPM) and Low Alarm (H2S: 10 PPM).These set points are factory set and cannot be changed. XAR-HP is equipped with audio alarms to alert you when the ambient gas concentration exceeds one of the two alarm set points. When H2S value exceeds the defined high alarm set point (15 PPM), the audio alarm will sound at 6 beeps/sec. When H2S value exceeds the defined low alarm set point (10 PPM) but less than the defined high alarm set point (15 PPM), the audio alarm will sound at 4 beeps/sec.

 

Factory Alarm Set points

Gas

Low

High

H2S

10 PPM

15 PPM

 

Specifications

H2S Sensor Specification:

Measurement Range

0 ~500 PPM

Resolution

1.0 PPM

Repeatability

<±5% of signal

Sample Method

Diffusion

Temperature Range

-20 to 40°C (-4°F ~ 104°F)

Humidity Range

15 to 90%RH

Response Time

<15 seconds

 

Temperature Specification

Temperature Range

-10.0~60.0°C (14~140°F) display

Display Resolution

0.1°C (0.1°F)

Display Options

°C/°F switchable

Accuracy

±0.5°C (±0.9°F)

Response Time

5~30 seconds (device must equilibrate with environment)

 

RH Specification

Measurement Range

0.0~99.9%RH

Display Resolution

1%RH

Accuracy

±4.5%RH

Response Time

<8 seconds for 63% of step change

 

General Specification

Operating

-10°C to40°C (14°F to 104°F),15~90% RH non-condensing

Storage

-10°C to 60°C (14°F to 140°F),<99% RH non-condensing

Power Supply

Four AA Alkaline Batteries

Dimensions

75x400x25mm (2.95x15.35x0.98”)

Weight

155 grams (5.47 oz.) without batteries