Portable Dissolved Oxygen Meter MHK-W30
MHK-W30 is a precise portable dissolved oxygen (DO) meter that measures dissolved oxygen in % and mg/L and temperature in water and is suitable for a wide range of applications, such as educational and aquaculture, as well as water and environmental analysis. It is suitable for monitoring the oxygen levels of lakes, shrimp and fish tanks.
Features
Measures dissolved oxygen in % and mg/L plus temperature
Automatic temperature compensation via temperature in probe provides accurate reading over the entire range
Adjustable barometric and salinity compensation
°C or °F unit switchable with 0.1°C/0.1°F resolution
Hold function freezes readings for easy viewing and recording
Support 1 and 2 points dissolved oxygen calibration at 100% water-saturated air and in 0 oxygen solution
Comes with a plastic toolbox convenient for in-field use
Powered by 3 AA alkaline batteries
Applications
Municipal wastewater/sewage treatment
Environmental water testing
Aquaculture/aquariums for measuring oxygen level in fish and shrimp farming
Natural stream purification system
Boilers
Educational in laboratories and schools
Biochemical manufacturing
Device
1. TFT display
2. Power button
3. Select button
4. Enter button
5. Dissolved oxygen polarographic electrode probe
6. Probe socket
7. Battery compartment cover
Display
1. Hold function prompt
2. Battery gauge
3. Dissolved oxygen value
4. Dissolved oxygen units (% or mg/L)
5. Barometric pressure (mmHg)
6. Salinity (in ppt)
7. Temperature
8. Temperature Unit
Barometric Pressure Compensation
The MHK-W30 uses a user inputted barometric pressure value in mmHg to compensate dissolved oxygen mg/L values. The barometric compensation value entered in the MHK-W30 should be the local barometric pressure value.
Do not use a barometer value that is corrected to sea level. Laboratory barometer readings are usually uncorrected values of air pressure and can be used for barometer calibration. An approximate formula for this “uncorrection” is:
True Barometric Pressure (mmHg) =
[Corrected BP mmHg] – [2.5 * (Local Altitude in ft. above sea level/100)]
Prior to the calibration or measurement, if true barometric pressure is unknown, users need to set the compatible parameter according to the local altitude. The following table 1 describes the relationship between altitude and barometric pressure.
Table 1 Relationship between Altitude and Barometric Pressure
Altitude (m) | mmHg | Altitude (m) | mmHg |
0 | 760 | 1600 | 622 |
100 | 750 | 1700 | 614 |
200 | 741 | 1800 | 607 |
300 | 732 | 1900 | 599 |
400 | 723 | 2000 | 592 |
500 | 714 | 2100 | 584 |
600 | 705 | 2200 | 577 |
700 | 696 | 2300 | 570 |
800 | 688 | 2400 | 563 |
900 | 679 | 2500 | 556 |
1000 | 671 | 2600 | 549 |
1100 | 662 | 2700 | 542 |
1200 | 654 | 2800 | 536 |
1300 | 646 | 2900 | 529 |
1400 | 638 | 3000 | 522 |
1500 | 630 | 3100 | 515 |
The barometric pressure can be adjusted any time without the need to recalibrate dissolved oxygen.
Salinity Compensation
The MHK-W30 uses a user inputted salinity value in ppt (parts per thousands) to compensate dissolved oxygen mg/L values. The salinity compensation value entered in the MHK-W30 should be the salinity value of the water you are testing.
The salinity can be adjusted any time without the need to recalibrate dissolved oxygen.
Specifications
Dissolved Oxygen Specification:
Range | 0 to 500% |
0 to 50.0 mg/L | |
Measuring principle | Dissolved oxygen polarographic electrode |
Resolution | 0.1 % |
0.1 mg/L | |
Accuracy | 0 to 200% air saturation :±2% of the reading or ± 2% air saturation, whichever is greater |
200 to 500% air saturation :±6% of the reading | |
0 to 20 mg/L :±2% of the reading or ±0.2 mg/L, whichever is greater | |
20 to 50 mg/L :±6% of the reading | |
Calibration Points | 1 to 2 points |
Barometric pressure compensation | 0 to 1000 mmHg |
Salinity compensation | 0 to 70.0 ppt |
Auto Temperature Compensation | 0 to 50.0 °C (32.0 to 122.0 °F) |
Temperature Specification:
Temperature Range | 0 to 50.0 °C (32.0 to 122.0 °F) |
Resolution | ± 0.1°C (± 0.1°F) |
Display Options | °C/°F switchable |
Accuracy | ±0.5°C (±0.9°F) |
General Specification:
Display | 2.2" TFT LCD |
Operating Temperature | 0 to 50.0 °C (32.0 to 122.0 °F) |
Power Supply | Three AA alkaline batteries |
Dimensions | 74x156x26.5mm (2.91x6.14x1.04") |
Weight | 125 grams (4.41 oz.) |